This panel includes:
Sodium (NAWB), potassium (KSB), chloride (CLWB) and ionized calcium (CAIB)
Iodide, bromide, thiosulfate, salicylate and nitrate cause a positive interference on chloride measured by ABL blood gas analyzers.
Ha, L et al, Annals Clin Biochem, 2015, 52(2): 288-292
Wendroth, S et al, Clin Chim Acta, 2014, 431: 77-79
ABL90 Flex Plus Instructions for Use Manual, Radiometer.
All reported values are corrected to 37C unless otherwise specified. Results beyond the linear range of the instrument will be reported as < or > the extreme of the linear range. Samples containing small bubbles may be run at the laboratory's discretion. If analyzed, a comment will beadded to the result regarding the presence of bubbles in the sample.
Plastic blood gas syringe containing approximately 70 IU dry heparin in a 3 mL syringe or 23 IU in a 1 mL syringe - or a capillary tube coated with 70 IU heparin (Radiometer Clinitube).
This panel includes:
Sodium (NAWB), potassium (KSB), chloride (CLWB) and ionized calcium (CAIB)
Iodide, bromide, thiosulfate, salicylate and nitrate cause a positive interference on chloride measured by ABL blood gas analyzers.
Ha, L et al, Annals Clin Biochem, 2015, 52(2): 288-292
Wendroth, S et al, Clin Chim Acta, 2014, 431: 77-79
ABL90 Flex Plus Instructions for Use Manual, Radiometer.
All reported values are corrected to 37C unless otherwise specified. Results beyond the linear range of the instrument will be reported as < or > the extreme of the linear range. Samples containing small bubbles may be run at the laboratory's discretion. If analyzed, a comment will beadded to the result regarding the presence of bubbles in the sample.
Plastic blood gas syringe containing approximately 70 IU dry heparin in a 3 mL syringe or 23 IU in a 1 mL syringe - or a capillary tube coated with 70 IU heparin (Radiometer Clinitube).
This panel includes:
Sodium (NAWB), potassium (KSB), chloride (CLWB) and ionized calcium (CAIB)
Iodide, bromide, thiosulfate, salicylate and nitrate cause a positive interference on chloride measured by ABL blood gas analyzers.
Ha, L et al, Annals Clin Biochem, 2015, 52(2): 288-292
Wendroth, S et al, Clin Chim Acta, 2014, 431: 77-79
ABL90 Flex Plus Instructions for Use Manual, Radiometer.
All reported values are corrected to 37C unless otherwise specified. Results beyond the linear range of the instrument will be reported as < or > the extreme of the linear range. Samples containing small bubbles may be run at the laboratory's discretion. If analyzed, a comment will beadded to the result regarding the presence of bubbles in the sample.
Ordering |
This panel includes:
Sodium (NAWB), potassium (KSB), chloride (CLWB) and ionized calcium (CAIB)
Iodide, bromide, thiosulfate, salicylate and nitrate cause a positive interference on chloride measured by ABL blood gas analyzers.
Ha, L et al, Annals Clin Biochem, 2015, 52(2): 288-292
Wendroth, S et al, Clin Chim Acta, 2014, 431: 77-79
ABL90 Flex Plus Instructions for Use Manual, Radiometer.
All reported values are corrected to 37C unless otherwise specified. Results beyond the linear range of the instrument will be reported as < or > the extreme of the linear range. Samples containing small bubbles may be run at the laboratory's discretion. If analyzed, a comment will beadded to the result regarding the presence of bubbles in the sample.
Collection |
Plastic blood gas syringe containing approximately 70 IU dry heparin in a 3 mL syringe or 23 IU in a 1 mL syringe - or a capillary tube coated with 70 IU heparin (Radiometer Clinitube).
Processing |
Result Interpretation |
This panel includes:
Sodium (NAWB), potassium (KSB), chloride (CLWB) and ionized calcium (CAIB)
Iodide, bromide, thiosulfate, salicylate and nitrate cause a positive interference on chloride measured by ABL blood gas analyzers.
Ha, L et al, Annals Clin Biochem, 2015, 52(2): 288-292
Wendroth, S et al, Clin Chim Acta, 2014, 431: 77-79
ABL90 Flex Plus Instructions for Use Manual, Radiometer.
All reported values are corrected to 37C unless otherwise specified. Results beyond the linear range of the instrument will be reported as < or > the extreme of the linear range. Samples containing small bubbles may be run at the laboratory's discretion. If analyzed, a comment will beadded to the result regarding the presence of bubbles in the sample.
Administrative |
Complete View |
Plastic blood gas syringe containing approximately 70 IU dry heparin in a 3 mL syringe or 23 IU in a 1 mL syringe - or a capillary tube coated with 70 IU heparin (Radiometer Clinitube).
This panel includes:
Sodium (NAWB), potassium (KSB), chloride (CLWB) and ionized calcium (CAIB)
Iodide, bromide, thiosulfate, salicylate and nitrate cause a positive interference on chloride measured by ABL blood gas analyzers.
Ha, L et al, Annals Clin Biochem, 2015, 52(2): 288-292
Wendroth, S et al, Clin Chim Acta, 2014, 431: 77-79
ABL90 Flex Plus Instructions for Use Manual, Radiometer.
All reported values are corrected to 37C unless otherwise specified. Results beyond the linear range of the instrument will be reported as < or > the extreme of the linear range. Samples containing small bubbles may be run at the laboratory's discretion. If analyzed, a comment will beadded to the result regarding the presence of bubbles in the sample.